Federal nursing home survey record · CMS-certified nursing home· Not a state assisted-living licensing record
Huntersville Health & Rehabilitation Center
Huntersville, NC · 90 certified beds · Last Life Safety survey May 7, 2025
CMS Certification Number 345570 · first certified June 2016
Ownership
Operated by LIFEWORKS REHAB · For profit - Corporation
No change of ownership on CMS record since January 1, 2016, when the records begin.
Position within NC
12 citations — more than 60% of the 419 certified nursing homes in NC. Compared within NC rather than nationally because which state a facility is in explains about 30% of the variation in citation counts between facilities, while bed count explains under 1%. State survey agencies differ far more than the buildings do. How this is calculated
The latest survey found 2 citations; the earlier surveys in the window averaged 5. With 3 surveys on file this is a direction of travel, not a conclusion.
When is the next survey likely?
Past the typical window (June 2026 to September 2026). Most NC facilities have been surveyed by September 2026. This facility’s last Life Safety survey was May 2025. Facilities in NC are typically surveyed 14–16 months after the last one (median 15), measured over 412 consecutive surveys in the last two years of CMS records.
A range the record supports, not a prediction of a date. State agencies vary widely — from about 11 months in Pennsylvania to three years in Maryland. How the window is measured
Get an email about Huntersville Health & Rehabilitation Center
One email when it happens. No account; stop it any time with one click.
Physical environment (health survey)
2 physical-environment citations on the health survey. Cited by the health surveyors under Appendix PP, not the Life Safety Code surveyor: equipment in safe operating condition, the call system, bed rails, pest control, hazards, the environment. Counted separately from the 12 Life Safety citations above. The Physical Environment Index
- F-0880 Provide and implement an infection prevention and control program.
How that compares
Compared with the median facility in NC, and nationally. Surveyors differ markedly between states, so the NC figure is the meaningful one.
| Measure | Citations |
|---|---|
| This facility | 12 |
| Median facility in NC | 9 |
| Median facility nationally | 11 |
Survey history
| Survey date | Citations |
|---|---|
| June 9, 2022 | 8 |
| January 25, 2024 | 2 |
| May 7, 2025 | 2 |
Most-cited tags
| Tag | What the surveyor checks | Times cited | Last cited |
|---|---|---|---|
| K-0918 | Have generator or other power source capable of supplying service within 10 seconds. | 1 | 2022-06-09 |
| K-0344 | Have an alternate power supply for its alarm system. | 1 | 2025-05-07 |
| K-0324 | Provide properly protected cooking facilities. | 1 | 2022-06-09 |
| K-0161 | Use approved construction type or materials. | 1 | 2025-05-07 |
| K-0353 | Inspect, test, and maintain automatic sprinkler systems. | 1 | 2022-06-09 |
| K-0100 | Meet other general requirements. | 1 | 2022-06-09 |
| K-0345 | Have approved installation, maintenance and testing program for fire alarm systems. | 1 | 2022-06-09 |
| K-0511 | Have properly installed electrical wiring and gas equipment. | 1 | 2022-06-09 |
What the citations cover
- Smoke Deficiencies 5
- Gas, Vacuum, and Electrical Systems Deficiencies 2
- Construction Deficiencies 2
- Services Deficiencies 2
- Other 1
| Category | Citations |
|---|---|
| Smoke Deficiencies | 5 |
| Gas, Vacuum, and Electrical Systems Deficiencies | 2 |
| Construction Deficiencies | 2 |
| Services Deficiencies | 2 |
| Egress Deficiencies | 1 |
Every citation on file
As published by CMS, newest survey first. Descriptions are CMS’s own wording.
May 7, 2025 — 2 citations
January 25, 2024 — 2 citations
| Tag | What the surveyor checks | Status |
|---|---|---|
| K-0223 | Provide exit doors that are held open by devices that will automatically close on the activation of a fire alarm or smoke detector. | Deficient, Provider has date of correction (March 22, 2024) |
| K-0923 | Have proper medical gas storage and administration areas. | Deficient, Provider has date of correction (March 22, 2024) |
June 9, 2022 — 8 citations
| Tag | What the surveyor checks | Status |
|---|---|---|
| K-0100 | Meet other general requirements. | Deficient, Provider has date of correction (September 20, 2022) |
| K-0324 | Provide properly protected cooking facilities. | Deficient, Provider has date of correction (September 20, 2022) |
| K-0345 | Have approved installation, maintenance and testing program for fire alarm systems. | Deficient, Provider has date of correction (September 20, 2022) |
| K-0353 | Inspect, test, and maintain automatic sprinkler systems. | Deficient, Provider has date of correction (September 20, 2022) |
| K-0363 | Install corridor and hallway doors that block smoke. | Deficient, Provider has date of correction (September 20, 2022) |
| K-0511 | Have properly installed electrical wiring and gas equipment. | Deficient, Provider has date of correction (September 20, 2022) |
| K-0521 | Ensure heating and ventilation systems that have been properly installed according to the manufacturer's instructions. | Deficient, Provider has date of correction (September 20, 2022) |
| K-0918 | Have generator or other power source capable of supplying service within 10 seconds. | Deficient, Provider has date of correction (September 20, 2022) |
Reflects the CMS posting of August 1, 2026. Surveys conducted in the last few months may not appear yet.