Federal nursing home survey record · CMS-certified nursing home· Not a state assisted-living licensing record

ECHELON CARE & REHAB

VOORHEES, NJ · 240 certified beds · Last Life Safety survey June 25, 2025

CMS Certification Number 315187 · first certified September 1981

Ownership

Operated by PARAMOUNT CARE CENTERS · For profit - Limited Liability company

  • Ownership changed May 10, 2022 (change of ownership)from THE LAKEWOOD OF VOORHEES OPERATOR
31
Citations on file
Rolling three-year window
3
Life Safety surveys
In the same window
8
Tags cited more than once
Across separate surveys
15
Inspection & testing records
Of the citations on file

Position within NJ

31 citations — more than 96% of the 348 certified nursing homes in NJ. Compared within NJ rather than nationally because which state a facility is in explains about 30% of the variation in citation counts between facilities, while bed count explains under 1%. State survey agencies differ far more than the buildings do. How this is calculated

Slippingmore citations at the latest survey than at its earlier surveys in the window.

The latest survey found 13 citations; the earlier surveys in the window averaged 9. With 3 surveys on file this is a direction of travel, not a conclusion.

When is the next survey likely?

The window opens October 2026. This facility’s last Life Safety survey was June 2025. Facilities in NJ are typically surveyed 16–20 months after the last one (median 17), measured over 431 consecutive surveys in the last two years of CMS records.

A range the record supports, not a prediction of a date. State agencies vary widely — from about 11 months in Pennsylvania to three years in Maryland. How the window is measured

Get an email about ECHELON CARE & REHAB

One email when it happens. No account; stop it any time with one click.

Which alerts

4 inspection-and-testing tags have been cited here at more than one survey. Nationally, that pattern is common: about three quarters of cited facilities have at least one repeated tag, and half of the facilities ever cited for sprinkler inspection and testing are cited for it again. A repeat is usually a recordkeeping gap rather than a new hazard.

Physical environment (health survey)

7 physical-environment citations on the health survey. Cited by the health surveyors under Appendix PP, not the Life Safety Code surveyor: equipment in safe operating condition, the call system, bed rails, pest control, hazards, the environment. Counted separately from the 31 Life Safety citations above. The Physical Environment Index

  • F-0584 Honor the resident's right to a safe, clean, comfortable and homelike environment, including but not limited to receiving treatment and supports for daily living safely.
  • F-0908 Keep all essential equipment working safely.
  • F-0921 Make sure that the nursing home area is safe, easy to use, clean and comfortable for residents, staff and the public.
  • F-0924 Put firmly secured handrails on each side of hallways.
  • F-0880 Provide and implement an infection prevention and control program.

How that compares

Citations on file over three years

Compared with the median facility in NJ, and nationally. Surveyors differ markedly between states, so the NJ figure is the meaningful one.

This facility31NJ median15National median11
Citations on file over three years, compared
MeasureCitations
This facility31
Median facility in NJ15
Median facility nationally11

Survey history

Citations at each Life Safety survey
92022-0192023-12132025-06
Citations at each Life Safety survey
Survey dateCitations
January 31, 20229
December 20, 20239
June 25, 202513

Most-cited tags

Most-cited tags at this facility
K-03633K-02813K-03512K-05212K-05412K-03532K-03742K-02912
Most-cited tags at this facility
TagWhat the surveyor checksTimes citedLast cited
K-0363Install corridor and hallway doors that block smoke.32025-06-25
K-0281Install proper backup exit lighting.32025-06-25
K-0351Install an approved automatic sprinkler system.22025-06-25
K-0521Ensure heating and ventilation systems that have been properly installed according to the manufacturer's instructions.22023-12-20
K-0541Install properly constructed and protected linen or trash chutes.22025-06-25
K-0353Inspect, test, and maintain automatic sprinkler systems.22025-06-25
K-0374Install smoke barrier doors that can resist smoke for at least 20 minutes.22023-12-20
K-0291Install emergency lighting that can last at least 1 1/2 hours.22025-06-25

What the citations cover

Citations by CMS category
  • Smoke Deficiencies 12
  • Egress Deficiencies 7
  • Gas, Vacuum, and Electrical Systems Deficiencies 6
  • Services Deficiencies 5
  • Other 1
Citations by CMS category
CategoryCitations
Smoke Deficiencies12
Egress Deficiencies7
Gas, Vacuum, and Electrical Systems Deficiencies6
Services Deficiencies5
Miscellaneous Deficiencies1

Every citation on file

As published by CMS, newest survey first. Descriptions are CMS’s own wording.

June 25, 2025 — 13 citations

Citations issued on June 25, 2025
TagWhat the surveyor checksStatus
K-0222Add doors in an exit area that do not require the use of a key from the exit side unless in case of special locking arrangements.Deficient, Provider has date of correction (July 31, 2025)
K-0271Have exits that are accessible at all times.Deficient, Provider has date of correction (July 31, 2025)
K-0281Install proper backup exit lighting.Deficient, Provider has date of correction (August 29, 2025)
K-0291Install emergency lighting that can last at least 1 1/2 hours.Deficient, Provider has date of correction (August 31, 2025)
K-0324Provide properly protected cooking facilities.Deficient, Provider has date of correction (June 20, 2025)
K-0351Install an approved automatic sprinkler system.Deficient, Provider has date of correction (August 31, 2025)
K-0353Inspect, test, and maintain automatic sprinkler systems.Deficient, Provider has date of correction (November 26, 2025)
K-0363Install corridor and hallway doors that block smoke.Deficient, Provider has date of correction (August 28, 2025)
K-0531Have elevators that firefighters can control in the event of a fire.Deficient, Provider has date of correction (July 23, 2025)
K-0541Install properly constructed and protected linen or trash chutes.Deficient, Provider has date of correction (August 12, 2025)
K-0761To conduct inspection, testing and maintenance of fire doors by qualified individuals.Deficient, Provider has date of correction (September 24, 2025)
K-0914Ensure receptacles at patient bed locations and where general anesthesia is administered, are tested after initial installation, replacement or servicing.Deficient, Provider has date of correction (July 22, 2025)
K-0921Ensure that testing and maintenance of electrical equipment is performed.Deficient, Provider has date of correction (July 28, 2025)

December 20, 2023 — 9 citations

Citations issued on December 20, 2023
TagWhat the surveyor checksStatus
K-0281Install proper backup exit lighting.Deficient, Provider has date of correction (February 16, 2024)
K-0351Install an approved automatic sprinkler system.Deficient, Provider has date of correction (February 16, 2024)
K-0355Properly select, install, inspect, or maintain portable fire extinguishes.Deficient, Provider has date of correction (February 9, 2024)
K-0363Install corridor and hallway doors that block smoke.Deficient, Provider has date of correction (March 7, 2024)
K-0374Install smoke barrier doors that can resist smoke for at least 20 minutes.Deficient, Provider has date of correction (February 9, 2024)
K-0521Ensure heating and ventilation systems that have been properly installed according to the manufacturer's instructions.Deficient, Provider has date of correction (February 9, 2024)
K-0541Install properly constructed and protected linen or trash chutes.Deficient, Provider has date of correction (March 22, 2024)
K-0911Meet requirements for the installation and maintenance of electrical systems.Deficient, Provider has date of correction (February 16, 2024)
K-0918Have generator or other power source capable of supplying service within 10 seconds.Deficient, Provider has date of correction (February 16, 2024)

January 31, 2022 — 9 citations

Citations issued on January 31, 2022
TagWhat the surveyor checksStatus
K-0281Install proper backup exit lighting.Deficient, Provider has date of correction (February 23, 2022)
K-0291Install emergency lighting that can last at least 1 1/2 hours.Deficient, Provider has date of correction (February 23, 2022)
K-0345Have approved installation, maintenance and testing program for fire alarm systems.Deficient, Provider has date of correction (February 23, 2022)
K-0353Inspect, test, and maintain automatic sprinkler systems.Deficient, Provider has date of correction (February 23, 2022)
K-0363Install corridor and hallway doors that block smoke.Deficient, Provider has date of correction (February 23, 2022)
K-0374Install smoke barrier doors that can resist smoke for at least 20 minutes.Deficient, Provider has date of correction (February 23, 2022)
K-0521Ensure heating and ventilation systems that have been properly installed according to the manufacturer's instructions.Deficient, Provider has date of correction (March 17, 2022)
K-0920Ensure proper usage of power strips and extension cords.Deficient, Provider has date of correction (February 23, 2022)
K-0923Have proper medical gas storage and administration areas.Deficient, Provider has date of correction (February 23, 2022)

Reflects the CMS posting of August 1, 2026. Surveys conducted in the last few months may not appear yet.