Federal nursing home survey record · CMS-certified nursing home· Not a state assisted-living licensing record
COMMUNITY CARE AND REHABILITATION CENTER
RIVERSIDE, CA · 162 certified beds · Last Life Safety survey May 9, 2025
CMS Certification Number 055409 · first certified January 1967
Ownership
Operated by DAVID JOHNSON · For profit - Limited Liability company
No change of ownership on CMS record since January 1, 2016, when the records begin.
Position within CA
24 citations — more than 67% of the 1,165 certified nursing homes in CA. Compared within CA rather than nationally because which state a facility is in explains about 30% of the variation in citation counts between facilities, while bed count explains under 1%. State survey agencies differ far more than the buildings do. How this is calculated
The latest survey found 4 citations; the earlier surveys in the window averaged 10. With 3 surveys on file this is a direction of travel, not a conclusion.
When is the next survey likely?
The window is open now: May 2026 to October 2026. This facility’s last Life Safety survey was May 2025. Facilities in CA are typically surveyed 12–18 months after the last one (median 14), measured over 1,766 consecutive surveys in the last two years of CMS records.
A range the record supports, not a prediction of a date. State agencies vary widely — from about 11 months in Pennsylvania to three years in Maryland. How the window is measured
Get an email about COMMUNITY CARE AND REHABILITATION CENTER
One email when it happens. No account; stop it any time with one click.
3 inspection-and-testing tags have been cited here at more than one survey. Nationally, that pattern is common: about three quarters of cited facilities have at least one repeated tag, and half of the facilities ever cited for sprinkler inspection and testing are cited for it again. A repeat is usually a recordkeeping gap rather than a new hazard.
Emergency preparedness
7 of the 24 citations on file are Emergency Preparedness E-tags. E-tags are cited under 42 CFR §483.73 and concern the emergency plan, training, drills, communications and supplies. Nationally about half of certified facilities have at least one. The Emergency Preparedness Index
Physical environment (health survey)
11 physical-environment citations on the health survey. Cited by the health surveyors under Appendix PP, not the Life Safety Code surveyor: equipment in safe operating condition, the call system, bed rails, pest control, hazards, the environment. Counted separately from the 24 Life Safety citations above. The Physical Environment Index
- F-0689 Ensure that a nursing home area is free from accident hazards and provides adequate supervision to prevent accidents.
- F-0880 Provide and implement an infection prevention and control program.
- F-0919 Make sure that a working call system is available in each resident's bathroom and bathing area.
- F-0584 Honor the resident's right to a safe, clean, comfortable and homelike environment, including but not limited to receiving treatment and supports for daily living safely.
How that compares
Compared with the median facility in CA, and nationally. Surveyors differ markedly between states, so the CA figure is the meaningful one.
| Measure | Citations |
|---|---|
| This facility | 24 |
| Median facility in CA | 19 |
| Median facility nationally | 11 |
Survey history
| Survey date | Citations |
|---|---|
| June 11, 2021 | 13 |
| April 25, 2024 | 7 |
| May 9, 2025 | 4 |
Most-cited tags
| Tag | What the surveyor checks | Times cited | Last cited |
|---|---|---|---|
| K-0363 | Install corridor and hallway doors that block smoke. | 3 | 2025-05-09 |
| K-0353 | Inspect, test, and maintain automatic sprinkler systems. | 2 | 2024-04-25 |
| K-0920 | Ensure proper usage of power strips and extension cords. | 2 | 2025-05-09 |
| E-0039 | Conduct testing and exercise requirements. | 2 | 2025-05-09 |
| E-0006 | Conduct risk assessment and an All-Hazards approach. | 2 | 2024-04-25 |
| K-0918 | Have generator or other power source capable of supplying service within 10 seconds. | 2 | 2024-04-25 |
| E-0032 | Provide primary/alternate means for communication. | 1 | 2024-04-25 |
| E-0041 | Implement emergency and standby power systems. | 1 | 2024-04-25 |
What the citations cover
- Smoke Deficiencies 10
- Emergency Preparedness Deficiencies 7
- Gas, Vacuum, and Electrical Systems Deficiencies 5
- Construction Deficiencies 1
- Other 1
| Category | Citations |
|---|---|
| Smoke Deficiencies | 10 |
| Emergency Preparedness Deficiencies | 7 |
| Gas, Vacuum, and Electrical Systems Deficiencies | 5 |
| Construction Deficiencies | 1 |
| Miscellaneous Deficiencies | 1 |
Every citation on file
As published by CMS, newest survey first. Descriptions are CMS’s own wording.
May 9, 2025 — 4 citations
| Tag | What the surveyor checks | Status |
|---|---|---|
| E-0039 | Conduct testing and exercise requirements. | Deficient, Provider has date of correction (May 26, 2025) |
| K-0161 | Use approved construction type or materials. | Deficient, Provider has date of correction (May 26, 2025) |
| K-0363 | Install corridor and hallway doors that block smoke. | Deficient, Provider has date of correction (May 26, 2025) |
| K-0920 | Ensure proper usage of power strips and extension cords. | Deficient, Provider has date of correction (May 26, 2025) |
April 25, 2024 — 7 citations
| Tag | What the surveyor checks | Status |
|---|---|---|
| E-0006 | Conduct risk assessment and an All-Hazards approach. | Deficient, Provider has date of correction (May 15, 2024) |
| E-0032 | Provide primary/alternate means for communication. | Deficient, Provider has date of correction (May 15, 2024) |
| E-0041 | Implement emergency and standby power systems. | Deficient, Provider has date of correction (May 15, 2024) |
| K-0345 | Have approved installation, maintenance and testing program for fire alarm systems. | Deficient, Provider has date of correction (May 15, 2024) |
| K-0353 | Inspect, test, and maintain automatic sprinkler systems. | Deficient, Provider has date of correction (May 15, 2024) |
| K-0363 | Install corridor and hallway doors that block smoke. | Deficient, Provider has date of correction (May 15, 2024) |
| K-0918 | Have generator or other power source capable of supplying service within 10 seconds. | Deficient, Provider has date of correction (May 15, 2024) |
June 11, 2021 — 13 citations
| Tag | What the surveyor checks | Status |
|---|---|---|
| E-0006 | Conduct risk assessment and an All-Hazards approach. | Deficient, Provider has date of correction (July 7, 2021) |
| E-0036 | Establish emergency prep training and testing. | Deficient, Provider has date of correction (July 7, 2021) |
| E-0039 | Conduct testing and exercise requirements. | Deficient, Provider has date of correction (July 7, 2021) |
| K-0324 | Provide properly protected cooking facilities. | Deficient, Provider has date of correction (July 7, 2021) |
| K-0343 | Have a fire alarm with audible and visual signals that transmits the alarm automatically to notify emergency forces in event of fire. | Deficient, Provider has date of correction (July 7, 2021) |
| K-0347 | Properly provide smoke detection systems in areas open to corridors. | Deficient, Provider has date of correction (July 7, 2021) |
| K-0351 | Install an approved automatic sprinkler system. | Deficient, Provider has date of correction (July 7, 2021) |
| K-0353 | Inspect, test, and maintain automatic sprinkler systems. | Deficient, Provider has date of correction (July 7, 2021) |
| K-0363 | Install corridor and hallway doors that block smoke. | Deficient, Provider has date of correction (July 7, 2021) |
| K-0761 | To conduct inspection, testing and maintenance of fire doors by qualified individuals. | Deficient, Provider has date of correction (July 7, 2021) |
| K-0914 | Ensure receptacles at patient bed locations and where general anesthesia is administered, are tested after initial installation, replacement or servicing. | Deficient, Provider has date of correction (July 7, 2021) |
| K-0918 | Have generator or other power source capable of supplying service within 10 seconds. | Deficient, Provider has date of correction (July 7, 2021) |
| K-0920 | Ensure proper usage of power strips and extension cords. | Deficient, Provider has date of correction (July 7, 2021) |
Reflects the CMS posting of August 1, 2026. Surveys conducted in the last few months may not appear yet.