Federal nursing home survey record · CMS-certified nursing home· Not a state assisted-living licensing record
AMS Memorial-Greene
Greene, IA · 31 certified beds · Last Life Safety survey July 16, 2026
CMS Certification Number 165356 · first certified November 1997
Ownership
Independently operated (no chain recorded by CMS) · For profit - Limited Liability company
- New ownershipOwnership changed June 1, 2025 (change of ownership)from MMS/GREENE HEALTHCARE LLC
Position within IA
41 citations — more than 100% of the 387 certified nursing homes in IA. Compared within IA rather than nationally because which state a facility is in explains about 30% of the variation in citation counts between facilities, while bed count explains under 1%. State survey agencies differ far more than the buildings do. How this is calculated
The latest survey found 24 citations; the earlier surveys on file averaged 8.5. With 3 surveys on file this is a direction of travel, not a conclusion.
When is the next survey likely?
The window opens June 2027 — about 7 months from now. This facility’s last Life Safety survey was July 2026. Facilities in IA are typically surveyed 11–13 months after the last one (median 12), measured over 692 consecutive surveys in the last two years of CMS records.
A range the record supports, not a prediction of a date. State agencies vary widely — from about 11 months in Pennsylvania to about three years in Virginia. How the window is measured
Get an email about AMS Memorial-Greene
One email when it happens. No account; stop it any time with one click.
5 inspection-and-testing tags have been cited here at more than one survey. Nationally, that pattern is common: about three quarters of cited facilities have at least one repeated tag, and half of the facilities cited for sprinkler inspection and testing in CMS’s current file are cited for it again. A repeat is usually a recordkeeping gap rather than a new hazard.
Emergency preparedness
9 of the 41 citations on file are Emergency Preparedness E-tags. E-tags are cited under 42 CFR §483.73 and concern the emergency plan, training, drills, communications and supplies. Nationally about half of certified facilities have at least one. The Emergency Preparedness Index
Physical environment (health survey)
2 physical-environment citations on the health survey. Cited by the health surveyors under Appendix PP, not the Life Safety Code surveyor: equipment in safe operating condition, the call system, bed rails, pest control, hazards, the environment. Counted separately from the 41 Life Safety citations above. The Physical Environment Index
How that compares
Compared with the median facility in IA, and nationally. Surveyors differ markedly between states, so the IA figure is the meaningful one.
| Measure | Citations |
|---|---|
| This facility | 41 |
| Median facility in IA | 12 |
| Median facility nationally | 11 |
Survey history
| Survey date | Citations |
|---|---|
| March 5, 2020 | 6 |
| February 21, 2022 | 11 |
| July 16, 2026 | 24 |
Most-cited tags
| Tag | What the surveyor checks | Times cited | Last cited |
|---|---|---|---|
| K-0372 | Ensure smoke barriers are constructed to a 1 hour fire resistance rating. | 2 | 2026-07-16 |
| K-0741 | Have posted "No-smoking" signs in areas where smoking is not permitted or ashtrays provided where smoking was allowed. | 2 | 2026-07-16 |
| K-0324 | Provide properly protected cooking facilities. | 2 | 2026-07-16 |
| K-0291 | Install emergency lighting that can last at least 1 1/2 hours. | 2 | 2026-07-16 |
| E-0004 | Develop and maintain an Emergency Preparedness Program (EP). | 2 | 2026-07-16 |
| K-0222 | Add doors in an exit area that do not require the use of a key from the exit side unless in case of special locking arrangements. | 2 | 2026-07-16 |
| K-0353 | Inspect, test, and maintain automatic sprinkler systems. | 2 | 2026-07-16 |
| K-0761 | To conduct inspection, testing and maintenance of fire doors by qualified individuals. | 2 | 2026-07-16 |
What the citations cover
- Smoke Deficiencies 14
- Emergency Preparedness Deficiencies 9
- Egress Deficiencies 8
- Miscellaneous Deficiencies 5
- Other 5
| Category | Citations |
|---|---|
| Smoke Deficiencies | 14 |
| Emergency Preparedness Deficiencies | 9 |
| Egress Deficiencies | 8 |
| Miscellaneous Deficiencies | 5 |
| Gas, Vacuum, and Electrical Systems Deficiencies | 4 |
| Services Deficiencies | 1 |
Every citation on file
As published by CMS, newest survey first. Descriptions are CMS’s own wording.
July 16, 2026 — 24 citations
| Tag | What the surveyor checks | Status |
|---|---|---|
| E-0004 | Develop and maintain an Emergency Preparedness Program (EP). | Deficient, Provider has date of correction (February 23, 2026) |
| E-0006 | Conduct risk assessment and an All-Hazards approach. | Deficient, Provider has date of correction (February 23, 2026) |
| E-0007 | Address patient/client population and determine types of services needed. | Deficient, Provider has date of correction (February 23, 2026) |
| E-0013 | Develop Emergency Preparedness policies and procedures. | Deficient, Provider has date of correction (February 23, 2026) |
| E-0029 | Develop a communication plan. | Deficient, Provider has date of correction (February 23, 2026) |
| E-0030 | List the names and contact information of those in the facility. | Deficient, Provider has date of correction (February 23, 2026) |
| E-0032 | Provide primary/alternate means for communication. | Deficient, Provider has date of correction (February 23, 2026) |
| E-0039 | Conduct testing and exercise requirements. | Deficient, Provider has date of correction (March 9, 2026) |
| K-0222 | Add doors in an exit area that do not require the use of a key from the exit side unless in case of special locking arrangements. | Deficient, Provider has date of correction (February 6, 2026) |
| K-0271 | Have exits that are accessible at all times. | Deficient, Provider has date of correction (January 13, 2026) |
| K-0281 | Install proper backup exit lighting. | Deficient, Provider has date of correction (January 21, 2026) |
| K-0291 | Install emergency lighting that can last at least 1 1/2 hours. | Deficient, Provider has date of correction (January 31, 2026) |
| K-0324 | Provide properly protected cooking facilities. | Deficient, Provider has date of correction (January 9, 2026) |
| K-0341 | Install a fire alarm system that can be heard throughout the facility. | Deficient, Provider has date of correction (January 16, 2026) |
| K-0345 | Have approved installation, maintenance and testing program for fire alarm systems. | Deficient, Provider has date of correction (January 7, 2026) |
| K-0346 | Follow proper procedures when the fire alarm was out of service for more than 4 hours. | Deficient, Provider has date of correction (February 23, 2026) |
| K-0347 | Properly provide smoke detection systems in areas open to corridors. | Deficient, Provider has date of correction (January 9, 2026) |
| K-0353 | Inspect, test, and maintain automatic sprinkler systems. | Deficient, Provider has date of correction (March 9, 2026) |
| K-0354 | Follow proper procedures when the automatic sprinkler systems was out of service for more than 10 hours. | Deficient, Provider has date of correction (February 23, 2026) |
| K-0372 | Ensure smoke barriers are constructed to a 1 hour fire resistance rating. | Deficient, Provider has plan of correction (July 5, 2026) |
| K-0741 | Have posted "No-smoking" signs in areas where smoking is not permitted or ashtrays provided where smoking was allowed. | Deficient, Provider has date of correction (January 7, 2026) |
| K-0761 | To conduct inspection, testing and maintenance of fire doors by qualified individuals. | Deficient, Provider has date of correction (January 8, 2026) |
| K-0918 | Have generator or other power source capable of supplying service within 10 seconds. | Deficient, Provider has date of correction (January 7, 2026) |
| K-0921 | Ensure that testing and maintenance of electrical equipment is performed. | Deficient, Provider has date of correction (February 5, 2026) |
February 21, 2022 — 11 citations
| Tag | What the surveyor checks | Status |
|---|---|---|
| E-0004 | Develop and maintain an Emergency Preparedness Program (EP). | Deficient, Provider has date of correction (February 25, 2022) |
| K-0222 | Add doors in an exit area that do not require the use of a key from the exit side unless in case of special locking arrangements. | Deficient, Provider has date of correction (March 24, 2022) |
| K-0223 | Provide exit doors that are held open by devices that will automatically close on the activation of a fire alarm or smoke detector. | Deficient, Provider has date of correction (March 24, 2022) |
| K-0291 | Install emergency lighting that can last at least 1 1/2 hours. | Deficient, Provider has date of correction (March 24, 2022) |
| K-0321 | Ensure that special areas are constructed so that walls can resist fire for one hour or have an approved fire extinguishing system. | Deficient, Provider has date of correction (March 24, 2022) |
| K-0324 | Provide properly protected cooking facilities. | Deficient, Provider has date of correction (March 22, 2022) |
| K-0355 | Properly select, install, inspect, or maintain portable fire extinguishes. | Deficient, Provider has date of correction (March 22, 2022) |
| K-0363 | Install corridor and hallway doors that block smoke. | Deficient, Provider has date of correction (March 24, 2022) |
| K-0741 | Have posted "No-smoking" signs in areas where smoking is not permitted or ashtrays provided where smoking was allowed. | Deficient, Provider has date of correction (March 3, 2022) |
| K-0920 | Ensure proper usage of power strips and extension cords. | Deficient, Provider has date of correction (February 24, 2022) |
| K-0923 | Have proper medical gas storage and administration areas. | Deficient, Provider has date of correction (February 24, 2022) |
March 5, 2020 — 6 citations
| Tag | What the surveyor checks | Status |
|---|---|---|
| K-0211 | Keep aisles, corridors, and exits free of obstruction in case of emergency. | Deficient, Provider has date of correction (March 11, 2020) |
| K-0353 | Inspect, test, and maintain automatic sprinkler systems. | Deficient, Provider has date of correction (March 11, 2020) |
| K-0372 | Ensure smoke barriers are constructed to a 1 hour fire resistance rating. | Deficient, Provider has date of correction (March 13, 2020) |
| K-0511 | Have properly installed electrical wiring and gas equipment. | Deficient, Provider has date of correction (March 13, 2020) |
| K-0712 | Have simulated fire drills held at unexpected times. | Deficient, Provider has date of correction (March 17, 2020) |
| K-0761 | To conduct inspection, testing and maintenance of fire doors by qualified individuals. | Deficient, Provider has date of correction (March 13, 2020) |
Reflects the CMS posting of October 1, 2026. Surveys conducted in the last few months may not appear yet.